Process Control Home
 
Alpha Controls
 

SmartReader Plus 7 Process Signal Data Logger

SmartReader Plus 7
Recording process signals is easy and convenient with the SmartReader Plus 7. With seven input channels it is ideal for monitoring (through commercially available transducers) a wide variety of measurement parameters. All process signals can be displayed in their own engineering units using the simple equation editor in ACR's TrendReader® Standard.


Eight Channel Process Signal Data Logger
SmartReader Plus 7 

Optional Input Ranges:

All external channels 0 to 2.5 volts, 0 to 5 volts, 0 to 10 volts, or 0 to 25mA

Accuracy:

+/– 0.5% of full scale

Memory Size:

32 KB, 128 KB and 1.5 MB

No. of Channels:

Eight. (One internal thermistor and 7 analog inputs)

Transducer Restrictions:

The use of a single power supply is recommended to avoid ground loops or use loop isolators

Overload Protection:

Voltage channels: +/-40 volts and Current channels: +/-70mA (reverse polarity protected)

Input type

0 to 2.5 volts DC channel 0 to 5.0 volts DC (2 channels) 0 to 10.0 volts DC channel 0 to 200 millivolts DC channel 0 to 25 milliamps DC (2 channels)

Input Impedance

Greater than 1M ohm Greater than 1M ohm Greater than 1M ohm Greater than 1M ohm 20 ohms

Remote Network Communications:

See Network & Communications

Industry:

Process Control, Process Signals, Commercial, Industrial, Production Characteristics Monitoring

PDF

SmartReader,Plus 7,Process,Signal,Data,Logger

Size: 107 x 74 x 22 mm (4.2" x 2.9" x 0.9")
Weight: 110 grams (3.75 ounces)
Case Material: NorylŪ Plastic
Operating Limits: -40°C to 70°C (-40°F to 158°F) and 0 to 95% Relative Humidity (non-condensing)
Clock Accuracy: +/-2 seconds per day
Battery: 3.6 volt Lithium, 1 Amp-Hour
Power Consumption: 5 to 10 microamps (continuous)
Battery Life: 10-years warrantied (under normal use). Factory replaceable
Memory Size: 32 K (21,500 readings). 128 K (87,000 readings). 1.5 MB (1,000,000 readings). See specific models for availability
Sampling Method: 1. Continuous (First-in, First-out); not available with sample rates faster than 8 seconds.) 2. Stop when full (Fill-then-stop)
Sampling Rate: User selectable rates from 25 per second to one reading every 8 hours. (BP-101 battery pack or PS-201 Power Supply required for sample rates faster than 8 seconds. See Accessories section for details)
Resolution: 12 bit (1 part in 4096)
PC Requirements: IBM PC or 100% compatible running MSŪ Windows 3.1,95,98, or NT, with a least 2 MB RAM, 2 MB of hard drive disk space and one free serial port
Product Approvals: Meets Part 15 for Digital Devices of the Code of Federal Regulations of the Federal Communications Commission (FCC). Meets Class A radiated and conducted emission requirements of Section 2 of the Radio Interference Regulations of Communications Canada. IEC801-2, 3, 4 and EN55011 covering ESD, RFI Immunity, EFT/ Burst and Radiated Emissions respectively for the Commission of the European Communities (CE)
Resistance to X-Rays: Tested for protection against a 160kV dose @ 5mA for 30 seconds (150mA-sec) @ 38 inches F.F.D. (about 100 times that of an airport X-Ray machine). Tested for protection against Gamma Ray (equivalent to 0.137-1.38 mega volts) IR 192 - 28 curies @ 30 second exposure source to object distance 5"


Powerful Solutions for Instrumentation & Control Problems!
SmartReader Plus 7 Process Signal Data Logger

Site Designed By Panther Marketing